• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SORIN GROUP ITALIA S.R.L. - CRM FACILITY PLATINIUM; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, CARDIAC RESYNCHRONIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SORIN GROUP ITALIA S.R.L. - CRM FACILITY PLATINIUM; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number PLATINIUM SONR CRT-D 1811
Device Problems No Device Output (1435); Over-Sensing (1438)
Patient Problem No Information (3190)
Event Date 06/26/2018
Event Type  malfunction  
Manufacturer Narrative
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by sorin that was cleared or approved by fda for marketing in the united states.
 
Event Description
Reportedly, one episode showing ventricular noise oversensing was observed in the device memory.Besides, the sonr signal amplitude significantly decreased since the replacement of the previous crt-d and tends towards zero since (b)(6) 2018.Preliminary analysis results confirmed the reported behavior.The observed ventricular noise most probably resulted from electromagnetic interferences (emi) and/or myopotentials.Based on available data, the origin of the loss of sonr signal could not be identified.
 
Manufacturer Narrative
Please refer to the attached analysis report.
 
Event Description
Reportedly, one episode showing ventricular noise oversensing was observed in the device memory.Besides, the sonr signal amplitude significantly decreased since the replacement of the previous crt-d and tends towards zero since mid (b)(6) 2018.Preliminary analysis results confirmed the reported behavior.The observed ventricular noise most probably resulted from electromagnetic interferences (emi) and/or myopotentials.Based on available data, the origin of the loss of sonr signal could not be identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINIUM
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
SORIN GROUP ITALIA S.R.L. - CRM FACILITY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR  92140
Manufacturer (Section G)
SORIN GROUP ITALIA SRL VIA CRESCENTINO 13040 SALUGGIA ITALY
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
Manufacturer Contact
elodie vincent
parc d'affaires noveos 4 avenue réaumur
.
clamart 92140
FR   92140
0146013665
MDR Report Key7705740
MDR Text Key114768792
Report Number1000165971-2018-00669
Device Sequence Number1
Product Code NIK
UDI-Device Identifier08031527014388
UDI-Public(01)08031527014388(11)161021(17)180521
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/21/2018
Device Model NumberPLATINIUM SONR CRT-D 1811
Device Catalogue NumberPLATINIUM SONR CRT-D 1811
Device Lot NumberS0226
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date06/27/2018
Date Manufacturer Received09/14/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-