• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE~RITE 8 ULTRASOUND SYSTEM (USA); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE~RITE 8 ULTRASOUND SYSTEM (USA); SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Model Number 9770500
Device Problem Poor Quality Image (1408)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will be evaluated.Results are expected soon.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
Rn reported ultrasound image on sr8 scanner looks grainy.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: during evaluation, the reported grainy ultrasound image issue could not be reproduced.The scanner was evaluated with the model 550 phantom (b)(4) and the image was found to be normal for a sr 8 scanner with the 32 mm probe.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.The review of the dhr, sub-assembly dhr, sub-assembly manufacturing, component records, manufacturing process changes, mrr/mrb, and internal rejects showed that the reported issue is not likely related to a manufacturing issue.No other functionality issues with the equipment were found during evaluation.The sr8 was serviced, qc tested, and returned to the customer.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
Rn reported ultrasound image on sr8 scanner looks grainy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 8 ULTRASOUND SYSTEM (USA)
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
Manufacturer (Section G)
DYMAX CORP. -2523003
110 marshall drive
warrendale PA 15086
Manufacturer Contact
sydney freckleton
605 n. 5600 w.
salt lake city, UT 84116
8015225992
MDR Report Key7711264
MDR Text Key114867432
Report Number3006260740-2018-01788
Device Sequence Number1
Product Code ITX
UDI-Device Identifier00801741096891
UDI-Public(01)00801741096891
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K152554
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number9770500
Device Catalogue Number9770500
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2018
Is the Reporter a Health Professional? Yes
Event Location Hospital
Date Manufacturer Received09/10/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-