No medical records or medical images have been made available to the manufacturer.As the lot number for the device was provided, a review of the device history records is currently being performed.The device is pending return to the manufacturer for evaluation.The investigation of the reported event is currently underway.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
Manufacturing review: the device history records have been reviewed with special attention to the raw materials, subassemblies, manufacturing process, and quality control testing.This lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the device was not returned for evaluation.Images and medical records were not provided for review.Therefore, the investigation is inconclusive for the alleged catheter malposition as no objective evidence has been provided to confirm any alleged deficiency with the catheter.The root cause could not be determined based upon available information.It is unknown whether patient and/or procedural factors contributed to the event.Possible contributing factors include improper placement; however, the lack of a returned sample prevented both confirmation of the reported event and identification of a root cause(s).Labeling review: a review of product labeling documents (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) showed that the product labeling is adequate.
|