Model Number 382400 |
Device Problems
No Audible Alarm (1019); No Display/Image (1183)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that an issue occurred with an enteral feeding pump.There was no patient involvement.Upon triage on (b)(6) 2018, service found the screen was blank when powered on by ac power and there were no start up tones.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
The unit was triaged and the reported conditions were confirmed.A trend has been identified for the no audible alarm condition and a corrective and preventative action has been opened to address this issue.For the screen/display issue, the potential root cause is customer misuse or general wear and tear.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.If information is provided in the future, a supplemental report will be issued.
|
|