• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVIS & GECK CARIBE LTD POLYSORB; ENDOSCOPIC TISSUE APPROXIMATION DEVICE Back to Search Results
Model Number 170051
Device Problems Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923); Human-Device Interface Problem (2949)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 06/25/2018
Event Type  malfunction  
Manufacturer Narrative
Occupation: health professional (hpro).If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, when the device was being applied to the jejunum during the sewing of the gastrojejunostomy on a laparoscopic roux en y gastric bypass, the device was difficult to unload at one point.The needle disengaged completely from the device but was still in the small bowel with the strand of suture attached, so it was visible the entire time.A new reload was used to complete the case.There was no patient injury.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
Post market vigilance (pmv) led an evaluation of a foil packaging and retainers of one device.Only foil packages and retainers were returned.Visual inspection found no abnormalities that would have caused or contributed to the reported condition.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all medtronic quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLYSORB
Type of Device
ENDOSCOPIC TISSUE APPROXIMATION DEVICE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO  0101
Manufacturer (Section G)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 0101
DO   0101
Manufacturer Contact
lisa hernandez
60 middletown ave
north haven, CT 06473
2034925563
MDR Report Key7712521
MDR Text Key114944137
Report Number9612501-2018-01338
Device Sequence Number1
Product Code OCW
UDI-Device Identifier20884521126753
UDI-Public20884521126753
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K934738
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 11/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2022
Device Model Number170051
Device Catalogue Number170051
Device Lot NumberJ7K1861X
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received10/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/19/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-