Model Number 1420 |
Device Problems
Connection Problem (2900); Material Twisted/Bent (2981); Positioning Problem (3009)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/18/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.Concomitant medical products: 6935m62 lead, dvfb1d4 icd implanted (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the data cable port of a patient¿s controller had a bent pin.It was noted that the patient's controller had not been dropped.The site further reported that they were unable to connect that port to the monitor data cable because it misaligned.The patient was reinstructed to use the data port cap to cover the data port on the controller when it¿s not in use.The controller was exchanged.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
The controller (con304423) was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.Visual inspection under 10x magnification revealed a hairline crack around power port 2.An internal visual inspection did not reveal evidence of fluid ingress.The observed hairline crack is not related to the reported event.Based on an investigation conducted, the root cause of the hairline crack was determined to be due to chemical additives applied to the power port gaskets during the manufacturing process.The chemical additives contributed to environmental stress cracking.Failure analysis also revealed a bent pin within the serial port of the controller.The bent pin did not allow a data cable to properly connect to the controller.As a result, the reported event was confirmed.The most likely root cause of the reported event can be attributed to a misalignment between the serial port and data cable.An internal investigation has been opened to evaluate bent/damaged pins with controller 2.0 and implement corrective actions as required.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
A supplemental report is being submitted for additional information.Additional information was received regarding the recall number.Investigation of this event is completed and the file will be closed.If new information is received, the file will be re-opened and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|