Model Number 990063-020
Device Problems Material Deformation (2976); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/04/2018
Event Type  malfunction  
Manufacturer Narrative
Product event summary: the data files and mapping catheter 990063-020 with lot number 214483806 were returned and analyzed.The data files confirmed system notice 50003"pressure is low in the system" at application 1 with a balloon catheter which was used for 1 injection.The data files also confirmed the system notice 50011 "obstructed flow" on the date of event.Visual inspection of mapping catheter showed that the shaft was kinked at 10 inches from the ecg connector.In conclusion, the mapping catheter failed the returned product inspection due to kink on the shaft.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The mapping catheter was returned and analyzed and it subsequently tested out of specification per the manufacturer's investigation.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7713738
MDR Text Key115196676
Report Number9612164-2018-01827
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/07/2019
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number214483806
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/07/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-