• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. MEDTRONIC LEAD; PULSE-GENERATOR, PACEMAKER, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. MEDTRONIC LEAD; PULSE-GENERATOR, PACEMAKER, IMPLANTABLE Back to Search Results
Model Number MDT-LEAD
Device Problems High impedance (1291); High Sensing Threshold (2574); High Capture Threshold (3266)
Patient Problems No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
Event Date 02/01/2017
Event Type  Injury  
Manufacturer Narrative
Initial reporter: non-healthcare professional.If information is provided in the future, a supplemental report will be issued.
 
Event Description
A law suit alleged that the implantable pulse generator (ipg) was suspected to have become non-functional.The right ventricular (rv) lead impedance and threshold values were high.It was also suspected that the ventricular lead was not connected at implant.The polarity of the lead was switched to unipolar.It was additionally noted that lead trend data showed the right atrial (ra) lead had high threshold.A re-operation procedure was planned.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product analysis: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated atrial pacing capture threshold was elevated.If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDTRONIC LEAD
Type of Device
PULSE-GENERATOR, PACEMAKER, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer (Section G)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7717776
MDR Text Key114946663
Report Number2182208-2018-01356
Device Sequence Number1
Product Code DXY
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,other
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMDT-LEAD
Device Catalogue NumberMDT-LEAD
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age75 YR
-
-