Model Number MDT-LEAD |
Device Problems
High impedance (1291); High Sensing Threshold (2574); High Capture Threshold (3266)
|
Patient Problems
No Known Impact Or Consequence To Patient (2692); No Code Available (3191)
|
Event Date 02/01/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Initial reporter: non-healthcare professional.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
A law suit alleged that the implantable pulse generator (ipg) was suspected to have become non-functional.The right ventricular (rv) lead impedance and threshold values were high.It was also suspected that the ventricular lead was not connected at implant.The polarity of the lead was switched to unipolar.It was additionally noted that lead trend data showed the right atrial (ra) lead had high threshold.A re-operation procedure was planned.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product analysis: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated atrial pacing capture threshold was elevated.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|