BOSTON SCIENTIFIC - MARLBOROUGH RESOLUTION CLIP CLIPPING DEVICE; CLIP, HEMOSTATIC Back to Search Results
Model Number M00522611
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/28/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.(b)(4).
 
Event Description
It was reported to boston scientific corporation that a resolution clip device was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the ¿long metal piece with the actual clip on it that opens and closes¿ fell inside the patient.Reportedly, the detached device was then removed, and the procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.Attempts to obtain additional information regarding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).Investigation results: a visual examination of the returned complaint device noted that the clip assembly was half deployed, as the yoke was still attached to the control wire.The clip assembly had an over bite and the clip prongs were locked onto the capsule windows.There were no visual issues noted with the bushing or along the catheter.This failure is likely due to anatomical/procedural factors encountered during the procedure which limited the performance of the device.Therefore, the most probable root cause is operational context.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.A search of the complaint database revealed that no similar complaints exist for the specified lot.
 
Event Description
It was reported to boston scientific corporation that a resolution clip device was used during a colonoscopy procedure performed on (b)(6) 2018.According to the complainant, during the procedure, the ¿long metal piece with the actual clip on it that opens and closes¿ fell inside the patient.Reportedly, the detached device was then removed, and the procedure was completed with another resolution clip device.There were no patient complications reported as a result of this event.The patient's condition at the conclusion of the procedure was reported to be fine.Attempts to obtain additional information regarding this event have been unsuccessful to date.Should additional relevant details become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTION CLIP CLIPPING DEVICE
Type of Device
CLIP, HEMOSTATIC
Manufacturer (Section D)
BOSTON SCIENTIFIC - MARLBOROUGH
100 boston scientific way
marlborough MA 01752
Manufacturer Contact
nancy cutino
100 boston scientific way
marlborough, MA 01752
5086834000
MDR Report Key7718314
MDR Text Key115070780
Report Number3005099803-2018-02443
Device Sequence Number1
Product Code MCH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040148
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/25/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/07/2020
Device Model NumberM00522611
Device Catalogue Number2261
Device Lot NumberML001443C5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/03/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age51 YR
Patient Weight100
-
-