Model Number 595000-001 |
Device Problem
Material Discolored (1170)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/16/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
This alleged failure mode poses a low risk to the patient because although the freedom driver's drivelines were discolored, the driver continued to perform its life-sustaining functions.The freedom driver has been returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.Ce: (b)(4).
|
|
Event Description
|
Upon receiving the freedom driver at syncardia, it was noted that its drivelines were discolored with a reddish/brown tint.
|
|
Manufacturer Narrative
|
The reported driveline discoloration was confirmed upon visual inspection of the drivelines.Root cause of the driveline discoloration could not be conclusively determined but was likely caused by exposure to a product in the patient's environment or a product that was applied topically to the drivelines.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
|
|