MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT¿; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT¿; STENT, CAROTID Back to Search Results
Model Number H7493915010310
Device Problem Capturing Problem (2891)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/28/2018
Event Type  malfunction  
Manufacturer Narrative
Device evaluated by mfr: the device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental medwatch will be filed.Bsc id: (b)(4).Tw: (b)(4).
 
Event Description
It was reported that stent reconstrainment difficulty occurred.A 10.0-31 carotid wallstent¿ was advanced to treat a lesion.Following deployment of about 20%, the physician decided to change the position.However, it was noted that resistance was met while resheathing and the device was retrieved while slightly deployed.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was good.
 
Manufacturer Narrative
Device evaluated by mfr: the delivery system was returned with the stent deployed.The deployed stent was not returned for analysis.A visual and tactile examination identified a severe shaft kinking and stretching damage along the length of the device.The majority of the damage was noted approximately 1260mm distal to the strain relief.Both the inner shaft and outer sheath were severely kinked and stretched.The outer sheath was also noted to be fully retracted and could not be repositioned over the inner shaft during analysis due to the severe shaft damage.The damage identified is consistent with excessive force being applied to the delivery system.No issues were noted with the catheter that could have contributed to the damage identified.A visual and microscopic examination was performed on the stent holder, stent cups and markerbands of the device.No damage or any issues were noted that could have contributed to the complaint incident.A visual and microscopic examination identified that the tip of the device was damaged.A section of the tip approximately 2mm in length was noted to be completely detached from the device.This detached section of tip was not returned for analysis.No other issues were identified during the product analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The investigation conclusion was unable to be determined.(b)(4).
 
Event Description
It was reported that stent reconstrainment difficulty occurred.A 10.0-31 carotid wallstent was advanced to treat a lesion.Following deployment of about 20%, the physician decided to change the position.However, it was noted that resistance was met while resheathing and the device was retrieved while slightly deployed.The procedure was completed with another of the same device.No patient complications were reported and the patient's status was good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT¿
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
MDR Report Key7723643
MDR Text Key115200985
Report Number2134265-2018-06798
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
PMA/PMN Number
P050019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 07/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/08/2021
Device Model NumberH7493915010310
Device Catalogue Number39150-1031
Device Lot Number0022103557
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2018
Date Manufacturer Received09/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-