BECTON, DICKINSON & CO. BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML; BLOOD SPECIMEN COLLECTION DEVICE Back to Search Results
Catalog Number 362780
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/05/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(6).A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.
 
Event Description
It was reported that a bd vacutainer® cpt¿ nh: ~130 iu ficoll¿: 2.0ml produced erroneous results.There was no report of exposure, serious injury or medical intervention.
 
Event Description
It was reported that a bd vacutainer cpt nh: ~130 iu ficoll: 2.0ml produced erroneous results.There was no report of exposure, serious injury or medical intervention.
 
Manufacturer Narrative
Due to an it issue beginning on (b)(6) 2018, previously filed emdrs did not contain required fields.This supplemental emdr is filed to provide the following omitted fields: event attributed to: other; device single use?: no; device returned to manufacture: no.Bd had not received samples, but photos were provided by the customer facility for investigation.The photos were evaluated and the customer's indicated failure mode for poor barrier separation with the incident lot was observed.A review of the device history record was completed for the incident lot and, based on this review, all product specifications and requirements for lot release were met; there were no related quality non-conformances during manufacturing of the product.Bd has initiated further investigation relating to this issue through a capa.The investigation is still on-going and improvements will be made as the potential causes of this issue are identified.Based on evaluation of the customer photos, the customer¿s indicated failure mode for poor barrier separation with the incident lot was observed.Further investigation has been initiated through a capa.The investigation is still on-going and improvements will be made as the potential causes are identified.A capa has been initiated to document further investigation and root cause analysis relating to this issue.The investigation is currently on-going and will be updated as the potential root cause(s) are identified.Based on an assessment of severity and frequency, it was determined that a capa is required at this time in order to determine the root cause associated with this issue and implement corrective actions.The investigation is currently on-going and further improvements will be made as the potential root cause(s) are identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD VACUTAINER® CPT¿ NH: ~130 IU FICOLL¿: 2.0ML
Type of Device
BLOOD SPECIMEN COLLECTION DEVICE
Manufacturer (Section D)
BECTON, DICKINSON & CO.
150 south 1st avenue
broken bow NE 68822
MDR Report Key7724650
MDR Text Key115365495
Report Number1917413-2018-03168
Device Sequence Number1
Product Code JCF
Combination Product (y/n)N
PMA/PMN Number
K891407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 09/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/26/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date11/30/2018
Device Catalogue Number362780
Device Lot Number7300904
Date Manufacturer Received07/05/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-