Model Number PM3210 |
Device Problems
Pacing Inadequately (1442); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/02/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Correction: the information previously submitted - initial reporter was incorrect.The information provided in this supplemental report is correct and should supersede the previously reported information.
|
|
Event Description
|
New information received indicated that the patient underwent an unrelated procedure during the episode.During the procedure, the pacemaker was temporarily reprogrammed to a low base rate which resulted in the long pacing delay episode.No further device intervention was performed.
|
|
Manufacturer Narrative
|
The results of the investigation are inconclusive since the device was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
|
Event Description
|
It was reported that the patient presented remotely via merlin.Net.Review of transmissions revealed that the pacemaker exhibited long pacing delays due to periodic switches in pacing rate.It was noted that the patient had their base rate programmed low during the episode.Patient was asymptomatic.No further information was reported.
|
|