• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS SITE~RITE 6 SCANNER; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS SITE~RITE 6 SCANNER; SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC Back to Search Results
Catalog Number 9770066
Device Problem Calibration Problem (2890)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will be evaluated.Results are expected soon.A history review of serial number (b)(4) showed no other similar complaint(s) from this serial number.
 
Event Description
Biomed reported end users complained of the sr6 ultrasound image being out of alignment, although the issue could not be confirmed when biomed was testing the system with tech support.
 
Event Description
Biomed reported end users complained of the sr6 ultrasound image being out of alignment, although the issue could not be confirmed when biomed was testing the system with tech support.
 
Manufacturer Narrative
The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of ultrasound out of alignment was unconfirmed.The needle tracking image is properly aligned.The root cause of the reported failure is inconclusive as the reported issue could not be reproduced during evaluation.The device was serviced, tested, and returned to the customer.A history review of serial number (b)(6) showed no other similar complaint(s) from this serial number.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SITE~RITE 6 SCANNER
Type of Device
SYSTEM, IMAGING, PULSED ECHO, ULTRASONIC
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key7729141
MDR Text Key115478023
Report Number3006260740-2018-01863
Device Sequence Number1
Product Code IYO
UDI-Device Identifier00801741091049
UDI-Public(01)00801741091049
Combination Product (y/n)N
PMA/PMN Number
K071204
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number9770066
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2018
Event Location Hospital
Date Manufacturer Received08/08/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-