• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORPORATION LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1; PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP Back to Search Results
Model Number 801041
Device Problem Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Per the user facility, the first time they started the pump it immediately stopped and displayed a "service pump" error message.Upon restarting the pump it displayed a "stopped: motor error" message.They disassembled the pump to check connections and clean the optical encoder.When the pump was reassembled and powered on it displayed a "stop: pump jam" error message.The distributor of the manufacturer's subsidiary worked with the manufacturer's technical support specialist to determine that the pump pod needed to be replaced.After replacing the pump pod the unit operated as intended with no error messages being displayed.
 
Event Description
It was reported that the roller pump displayed multiple error messages.No other details regarding the nature of the event were provided.
 
Manufacturer Narrative
The reported complaint was confirmed.The unit was not returned so there was not testing conducted by the manufacturer.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LARGE (6) ROLLER PUMP FOR TERUMO ADVANCED PERFUSION SYSTEM 1
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, ROLLER TYPE ¿ 6 INCH ROLLER PUMP
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7732259
MDR Text Key115525665
Report Number1828100-2018-00393
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
K112587
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/30/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801041
Device Catalogue Number801041
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/12/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/24/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-