Model Number 39304 |
Device Problems
Nonstandard Device (1420); Appropriate Term/Code Not Available (3191)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
|
|
Event Description
|
It was reported that the adhesive of the catheter was too strong.
|
|
Manufacturer Narrative
|
The device was not returned for evaluation.The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.".
|
|
Event Description
|
It was reported that the adhesive of the catheter was too strong.
|
|
Search Alerts/Recalls
|