Model Number 990063-020
Device Problems Material Deformation (2976); Output Problem (3005)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/29/2018
Event Type  malfunction  
Manufacturer Narrative
Non-healthcare professional product event summary: the data files and mapping catheter, (b)(4) were returned and analyzed.The data files showed 6 injections were performed with a balloon catheter and showed system notice (b)(4) ¿fluid in the catheter¿ at the final injection.Also, 2 injections were performed with a different catheter without any issues or system notices.Visual inspection of the mapping catheter showed the loop had some kinks.The mapping catheter failed the return product inspection.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The mapping catheter was returned and subsequently tested out of specification per the manufacturer's investigation.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACHIEVE MAPPING CATHETER
Type of Device
CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
Manufacturer (Section D)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX  22570
Manufacturer (Section G)
MEDTRONIC MEXICO
av. paseo del cucapah #10510
tijuana,bc 22570
MX   22570
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7739260
MDR Text Key115778774
Report Number9612164-2018-01923
Device Sequence Number1
Product Code DRF
Combination Product (y/n)N
Reporter Country CodeMG
PMA/PMN Number
K102588
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/01/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2019
Device Model Number990063-020
Device Catalogue Number990063-020
Device Lot Number213478293
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/24/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/14/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age78 YR
Patient Weight135
-
-