PRO MED INSTRUMENTS GMBH DORO® SKULL CLAMP Back to Search Results
Model Number 1001.001
Device Problems Use of Device Problem (1670); Device Handling Problem (3265)
Patient Problem Laceration(s) (1946)
Event Type  Injury  
Manufacturer Narrative
The device was inspected and tested on (b)(4) 2018.Within this inspection functional testing and visual inspection was made.No deviation was found.We suspect, that maybe the pinning technique has been not optimal as described in the instruction manual: "adjust the skull clamp to the width of the patient's head in the manner that the two skull pins in the rocker arm are equidistant from the centerline of the head and the single skull pin at the extension assembly is in line with this centerline.".
 
Event Description
Customer service was contacted on (b)(6) 2018 by customer.Customer stating that this item needs inspection and evaluation.On (b)(6), the customer stated that this skull clamp was involved in a slippage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DORO® SKULL CLAMP
Type of Device
DORO® SKULL CLAMP
Manufacturer (Section D)
PRO MED INSTRUMENTS GMBH
boetzinger str. 38
freiburg, baden-wuerttemberg 79111
GM  79111
Manufacturer (Section G)
PRO MED INSTRUMENTS GMBH
boetzinger str. 38
freiburg, baden-wuerttemberg 79111
GM   79111
Manufacturer Contact
sandra untenberger
boetzinger str. 38
freiburg, baden-wuerttemberg 79111
GM   79111
MDR Report Key7741775
MDR Text Key115764037
Report Number3003923584-2018-00039
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001808
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other
Type of Report Initial
Report Date 07/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/02/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number1001.001
Device Catalogue Number1001.001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/06/2018
Date Manufacturer Received07/06/2018
Was Device Evaluated by Manufacturer? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-