Model Number 36102
Device Problems Nonstandard Device (1420); Output above Specifications (1432)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that the adhesive peel strength for the samples returned were above the specification for strong adhesive.
 
Manufacturer Narrative
The reported event was confirmed as manufacturer related.Per the evaluation, two of the five tested samples were found above specification.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) open package at perforation.4) to remove plastic insert, squeeze catheter at the top of the white cone and pull to release.5) unroll self-adhering catheter over penis.6) gently squeeze the catheter to properly seal adhesive to the skin.7) connect to collection bag.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive.".
 
Event Description
It was reported that the adhesive peel strength for the samples returned were above the specification for strong adhesive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC WIDEBAND M 100BX
Type of Device
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7746897
MDR Text Key116121265
Report Number1018233-2018-03293
Device Sequence Number0
Product Code NNX
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,other
Type of Report Initial,Followup
Report Date 09/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/03/2018
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date09/28/2022
Device Model Number36102
Device Catalogue Number36102
Device Lot NumberJUBX0378
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2018
Date Manufacturer Received08/13/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-