The product was returned for evaluation.The returned unit passed all specific functional testing requirements.When the unit was properly positioned and put under pressure, it functioned properly.The product needs heli-coils added to large starburst threads, general maintenance and cleaning is required.The device history record for a1059 was reviewed and there were no issues found on the build record.The complaint was not confirmed.The root cause cannot be determined at this time.
|
A customer reported that on (b)(6) 2018, an a1059 mayfield modified skull clamp was used in an unspecified procedure.It was reported that the screw part on the end was not gauging to the correct number once on the patient.Additional information was received on 11jul2018 stating that the device was used on a (b)(6) male patient.There was surgery delay of around 5-10 minutes, just long enough to get the other c-clamp to complete the surgery.There was no patient injury and no medical revision required.
|