MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/09/2018
Event Type  malfunction  
Manufacturer Narrative
The product was returned for evaluation.The returned unit passed all specific functional testing requirements.When the unit was properly positioned and put under pressure, it functioned properly.The product needs heli-coils added to large starburst threads, general maintenance and cleaning is required.The device history record for a1059 was reviewed and there were no issues found on the build record.The complaint was not confirmed.The root cause cannot be determined at this time.
 
Event Description
A customer reported that on (b)(6) 2018, an a1059 mayfield modified skull clamp was used in an unspecified procedure.It was reported that the screw part on the end was not gauging to the correct number once on the patient.Additional information was received on 11jul2018 stating that the device was used on a (b)(6) male patient.There was surgery delay of around 5-10 minutes, just long enough to get the other c-clamp to complete the surgery.There was no patient injury and no medical revision required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH 45227
Manufacturer Contact
kimberly shelly
311 enterprise drive
plainsboro, NJ 08536
6099362393
MDR Report Key7753773
MDR Text Key116256307
Report Number3004608878-2018-00141
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 07/10/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number139
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/09/2018
Date Manufacturer Received07/10/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/27/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age77 YR
-
-