• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PLEXUS MANUFACTURING SDN. BHD; PULSE-GENERATOR, PACEMAKER, EXTERNAL Back to Search Results
Model Number 53401
Device Problems Electrical /Electronic Property Problem (1198); Mechanical Problem (1384); Nonstandard Device (1420); Failure to Read Input Signal (1581)
Patient Problem No Patient Involvement (2645)
Event Date 07/24/2018
Event Type  malfunction  
Manufacturer Narrative
Analysis found that the output dial failed.As a result the encoder assembly was replaced.A firmware update was also performed.The device then passed all final functional testing.This device was reported as included in the field action noted but returned product investigation found the device did not perform as described in the field action.The device is no longer included as part of the field action.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a reset to nominal values could be induced on the external pulse generator (epg) by hitting, rubbing or shaking the device.This was noticed only during testing of the device by the technical department in the hospital.The device has been quarantined by the hospital and will not be returned.There was no patient involvement.The device was later returned for a firmware update.It was analyzed and tested out of specification.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
PULSE-GENERATOR, PACEMAKER, EXTERNAL
Manufacturer (Section D)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY  11900
Manufacturer (Section G)
PLEXUS MANUFACTURING SDN. BHD
bayan lepas free industrial zo
bayan lepas 11900
MY   11900
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key7753946
MDR Text Key116609592
Report Number3004593495-2018-00718
Device Sequence Number1
Product Code DTE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 10/04/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/06/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number53401
Device Catalogue Number53401
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/26/2018
Is the Reporter a Health Professional? No
Date Manufacturer Received08/08/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2317-2018
Patient Sequence Number1
-
-