During preparation for a thrombectomy procedure, the hospital found that the penumbra system ace 60 reperfusion catheter (ace60) could not be flushed in a proper way for unknown reasons.The reported issue was found prior to use.Therefore, the ace60 was not used in the procedure.The procedure was completed using a new ace60.
|
Results: there was no visible damage to the ace60.The distal tip of the ace60 had coagulated blood inside.Conclusions: evaluation of the returned ace60 revealed no visible damage.During decontamination and functional test, the ace60 was flushed using the decontamination solution and water without issue.Further evaluation of the returned ace60 revealed that the distal tip of the catheter had coagulated blood inside.The coagulated blood likely contributed to the inability to flush the ace60 during the procedure.Although the complaint stated that the device was not used in the procedure, blood occluded the tip of the returned device.This indicates the ace60 was likely used in the procedure.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|