Model Number RD900 |
Device Problem
Device Handling Problem (3265)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/06/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The complaint rd900 neopuff infant resuscitator is currently enroute to fisher & paykel healthcare in (b)(4) for evaluation, to determine if it caused or contributed to the reported event.We will provide a follow up report upon completion of our investigation.
|
|
Event Description
|
A healthcare facility in (b)(6) reported that a rd900 neopuff infant resuscitator unit is losing pressure.This issue was noticed during a preventative maintenance check and requested service.There was no patient involvement.
|
|
Event Description
|
A healthcare facility in iowa reported that a rd900 neopuff infant resuscitator unit is losing pressure.This issue was noticed during a preventative maintenance check and requested service.There was no patient involvement.
|
|
Manufacturer Narrative
|
(b)(4).Method: only the manometer of the complaint rd900 neopuff infant resuscitator was returned to fisher & paykel healthcare for investigation.No physical damage observed but the manometer was received with the needle indicating approximately -12cmh2o.The complaint manometer was adjusted to indicate 0cmh2o and fitted into a known good valve system and tested based on the neopuff technical manual.Results: the performance test revealed that the manometer of the subject neopuff was out of specification.Conclusion: it is most likely that the out of specification manometer was caused by some sort of impact to the neopuff.The neopuff is a portable, reusable device used to assist in the delivery of respiratory breaths to infants until adequate spontaneous breathing occurs.Being a portable device, the neopuff can be susceptible to impact damage, for instance when accidentally dropped or subjected to considerable external force.It should be noted that the complaint device was more than twelve years old.The neopuff technical manual states the following: "dropping the neopuff / perivent infant resuscitator or other similar forms of impact may cause damage resulting in incorrect operation of the unit.If you suspect damage to have occurred, please perform checks as outlined [in the manual] before connection to a patient".In addition the neopuff user instructions state that the user should "check manometer reads zero with no gas flow" and check the pressure settings "prior to every use of the neopuff".The subject neopuff unit was repaired and returned to the customer after passing all the necessary performance tests.
|
|
Search Alerts/Recalls
|