Model Number 801763
Device Problems Overheating of Device (1437); Electrical Shorting (2926)
Patient Problem No Patient Involvement (2645)
Event Date 07/18/2018
Event Type  malfunction  
Manufacturer Narrative
The pst observed the integrated circuit (ic) u16 to be blown and a hole in the plastic package near pins six and seven on the power manager board, which is consistent with of a catastrophic power event such as an inadvertent short circuit.There was also charring of the plastic case of ic u13 observed near pins one and two on the network interface card (nic), which are ground and +5 pins respectively.The power manager board and both network nic boards attached to the +5 vdc and gnd on all three boards were affected.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Event Description
The product surveillance technician (pst) reported that during routine testing of the device at the service center, the five volts direct current (vdc) to ground (gnd) had a short circuit.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED PERFUSION SYSTEM 1
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS-ADVANCED PERFUSION SYSTEM 1
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
6200 jackson road
ann arbor MI 48103
Manufacturer (Section G)
SAME
Manufacturer Contact
douglas patton
6200 jackson road
ann arbor, MI 48103
7346634145
MDR Report Key7767943
MDR Text Key116709443
Report Number1828100-2018-00416
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K172220
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number801763
Device Catalogue Number801763
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/04/2003
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-