Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied with eight clips remaining.No visual abnormalities were observed with the instrument.The instrument was applied to appropriate test media for functional evaluation.The instrument was found to cycle without binding.Clips advanced into the jaws, formed properly, and were held securely in place after full formation was achieved and the firing handle was released.In addition, when the cartridge was empty, the interlock engaged to prevent the jaws from approximating.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.If information is provided in the future, a supplemental report will be issued.
|