• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; NRY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SYSTEM ASPIRATION PUMP MAX 110; NRY Back to Search Results
Catalog Number PMX110
Device Problems Misconnection (1399); Device Contamination with Body Fluid (2317); Device Handling Problem (3265)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/12/2018
Event Type  malfunction  
Manufacturer Narrative
Results: dried blood was present on the pump max device housing.A pipe cleaner was inserted into the vacuum pump and blood was present.Conclusions: evaluation of the returned pump max revealed blood on the device housing and inside the vacuum pump.If the aspiration tubing is connected directly to the vacuum inlet instead of to the penumbra supplied canister, blood will likely enter the vacuum pump.Penumbra pumps are visually inspected and functionally tested during incoming quality inspection.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Event Description
The patient was undergoing a thrombectomy procedure using a penumbra system aspiration pump max 110v (pump max).During the procedure, the hospital technologist accidentally connected the aspiration tubing (tubing) directly to the pump max rather than the canister.Consequently, blood was aspirated into the pump max during the first pass.Upon realizing the mistake, the tubing was reconnected to the canister.The procedure was successfully completed using the same pump max.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SYSTEM ASPIRATION PUMP MAX 110
Type of Device
NRY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
Manufacturer Contact
veronica farris
one penumbra place
alameda, CA 94502
5107483200
MDR Report Key7770049
MDR Text Key116717734
Report Number3005168196-2018-01551
Device Sequence Number1
Product Code NRY
UDI-Device Identifier00814548012773
UDI-Public00814548012773
Combination Product (y/n)Y
Reporter Country CodeUS
PMA/PMN Number
K160449
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/01/2005,07/12/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberPMX110
Device Lot NumberF07792-14
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/31/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received07/12/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age18 YR
-
-