• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MEDICAL PRODUCTS KENDALL SCD; SLEEVE, LIMB, COMPRESSIBLE Back to Search Results
Model Number 295251
Device Problems Disconnection (1171); Power Problem (3010)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Customer states that the power cord plug is broken.Upon triage it was noted that the live pin on the plug is broken.
 
Manufacturer Narrative
An investigation was performed for the reported customer complaint: ¿customer states that the power cord plug is broken.Upon triage it was noted that the live pin on the plug is broken.¿ a review of the device history record (dhr) for lot no.V1121056sx indicated the product was released meeting all quality standards.All dhrs are reviewed for accuracy prior to product release.In-process procedures are also in place to prevent nonconforming product in the manufacturing process.This ensures components and finished products meet all quality inspection standards.These controls include, but are not limited to: material verification/certification processes, dimensional specifications, statistical samplings, periodic audits, process inspections, machine maintenance/operation and personnel training and certification.The unit was returned for preventative maintenance and evaluated for the reported customer complaint.Visual inspection observed the bed hook and the plug of the power cord were broken.The bed hook and power cord were replaced.After component replacement, the device was fully tested for functionality and met manufacturer's guidelines.In addition, electrical safety testing was performed pursuant to iec62353 with passing results.The reported customer complaint is confirmed.The root cause is determined to be customer misuse.This complaint will be used for tr ending purposes.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KENDALL SCD
Type of Device
SLEEVE, LIMB, COMPRESSIBLE
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
MDR Report Key7771981
MDR Text Key116721928
Report Number3006451981-2018-00562
Device Sequence Number1
Product Code JOW
UDI-Device Identifier10884521129580
UDI-Public10884521129580
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 11/05/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number295251
Device Catalogue Number295251
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2018
Date Manufacturer Received08/09/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-