MEDTRONIC SINGAPORE OPERATIONS ADVISA DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number A3DR01 |
Device Problems
Mechanical Problem (1384); Material Integrity Problem (2978)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 05/21/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the physician put the right ventricular (rv) lead in and rotated the set screw clockwise but the lead would not fixate, and the torque wrench did not work.The physician tried another wrench with the same result.The lead was never implanted, and a new lead was used.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.If information is provided in the future, a supplemental report will be issued.
|
|