Catalog Number 228143
Device Problems Activation, Positioning or Separation Problem (2906); Activation Problem (4042)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/20/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date is not currently available.Associated medwatches: 1221934-2018-54470, 1221934-2018-54471, 1221934-2018-54472.
 
Event Description
It was reported during shoulder arthroscopy procedure on (b)(6) 2018, second implant could not be fired.The depth stop restricted the opening for the second implant.They opened a second omnispan gun which had the same problem.Only the third omnispan worked.The surgery took more time, they had to open a new omnispan gun and it frustrated the surgeon.
 
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate evaluation statement: the complaint device was received and inspected visually and functionally.Visual observation revealed only a slight bend in the pusher rod on the device.When tested for its functionality, both triggers functioned properly when pulled on their own without a needle attached to the device.Additionally, a 0 degree needle was loaded onto the applier and both implants were deployed successfully.The applier works as intended and we cannot confirm the reported complaint.Per qlik query executed, no nonconformances were identified for this part-lot number combination.Furthermore, a review into the depuy synthes mitek complaints system revealed one similar complaint for this lot of devices.At this point in time, no corrective action is required and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).Associated medwatch report number: 1221934-2018-54470, 1221934-2018-54471, 1221934-2018-54472.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENISCAL DEPLOYMENT GUN
Type of Device
ORTHOPAEDIC CERCLAGE APPLIER
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
jennifer lawrence
325 paramount drive
raynham, MA 02767
5088808100
MDR Report Key7774960
MDR Text Key117153597
Report Number1221934-2018-54469
Device Sequence Number1
Product Code GEF
UDI-Device Identifier10886705010059
UDI-Public10886705010059
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 07/13/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/10/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number228143
Device Lot NumberL513127
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/02/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/10/2018
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-