Model Number NOT APPLICABLE
Device Problem Obstruction of Flow (2423)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/19/2018
Event Type  malfunction  
Manufacturer Narrative
The system was used for treatment.This case is reportable as a mdr due to the reportable malfunction clot observed in the return line.Since this reportable malfunction is only associated with the kit, this mdr will only be against the kit.A batch record review for kit lot g312 was conducted.There were no non-conformances associated with this lot.This lot met all release requirements.A review of kit lot g312 for the reported issue shows no trends.Trends were reviewed for complaint categories, alarm #1: air detected and clot observed.No trends were detected for each complaint category.The assessment is based on the information available at the time of investigation.No product was returned for investigation; therefore, it could not be determined if the product met specification based solely on the information provided by the customer.Complaints are monitored through tracking and trending.If a trend is detected, further investigation will be conducted.Investigation complete.(b)(4).
 
Event Description
The customer called to report a blood clot was observed during the treatment procedure.The customer stated at approximately 284 ml of whole blood processed they received an alarm #1: air detected alarm and noted air in the return line.The customer stated they completed a saline bolus to clear the air from the line and saw blood clots coming out of the line.The customer aborted the procedure and did not return blood to the patient.The customer reported the patient was stable.The customer opened the centrifuge door to remove the kit and found clotting in the centrifuge bowl.The customer indicated that heparin was administered at an 8:1 ratio and the heparin bag was correctly spiked.The customer will not be returning product for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELLEX PHOTOPHERESIS SYSTEM
Type of Device
CELLEX PHOTOPHERESIS SYSTEM
Manufacturer (Section D)
THERAKOS, INC.
bedminster NJ 07921
Manufacturer (Section G)
MALLINCKRODT PHARMACEUTICALS IRELAND LIMITED
college business & tech park
cruiserath road
blanchardstown, dublin D15 T X2V
EI   D15 TX2V
Manufacturer Contact
megan vernak
1425 us route 206
bedminster, NJ 07921
MDR Report Key7787725
MDR Text Key117450545
Report Number2523595-2018-00129
Device Sequence Number1
Product Code LNR
UDI-Device Identifier20705030200003
UDI-Public(01)20705030200003(10)G312(17)200201
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Nurse
Type of Report Initial
Report Date 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/01/2020
Device Model NumberNOT APPLICABLE
Device Catalogue NumberCLXUSA
Device Lot NumberG312
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/14/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-