Catalog Number BI70002000-G27
Device Problem Difficult to Open or Close (2921)
Patient Problem No Patient Involvement (2645)
Event Date 07/23/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.It was noted that a non-medtronic personnel made repairs to the system.The system is operational.
 
Event Description
Medtronic received information regarding an imaging system being used outside of a procedure.It was reported that the system was stating that the gantry is open even when they completely close it.Multiple reboots did not resolve the issue as well as trying to open the gantry completely then re-closing it.There was no patient present.
 
Manufacturer Narrative
Manufacturer updated to proper value.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
peter verhey
navigation customer quality
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7789430
MDR Text Key117434244
Report Number1723170-2018-04087
Device Sequence Number1
Product Code OWB
UDI-Device Identifier00643169639683
UDI-Public00643169639683
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 09/20/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBI70002000-G27
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/17/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-