• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY AG ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY AG ROTAFLOW CENTRIFUGAL PUMP SYSTEM; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number ROTAFLOW CONSOLE
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 07/25/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.Reference exemption # (b)(4).Importer- (b)(4).The issue was observed at the (b)(4) depot.According to the service report# (b)(4) dated on (b)(4) 2018 following work has been performed: the rotalow console displayed head error when rpm is increased above 2500 rpm ans is non-functional.The service technician replaced the rf control board pcba kit (material# (b)(4), serial# (b)(4)).Preventive maintenance, functional test ans safety check as per the service manual have been performed.The rotaflow console passed all tests.Thus the failure could be confirmed.In the course of the investigation of (b)(4) all complaints were analyzed individually to look for patterns and causes.After evaluation of the complaints, the following defects are the most common: hot plug, sig error followed by head error, error message due to shaking / error message due to sensitivity, connection problems and hardware-error.The increase in complaints with the error "head error" is due to a user / application error.The actions have been addressed in the past to prevent application errors.Furthermore, the instructions for use of the rotaflow system, see rotaflow system user manual, (b)(4), contain detailed descriptions to prevent an "error head".In addition, the instruction manual describes how the user has to react in case of an error message so that the application can be continued if possible.Since there are several causes of errors that result in an error head (error head) message, no final root cause could be determined.It is noted that the evaluated error is largely due to a user error.Warning in the ifu regarding "hotplug" and label with the "hotplug" warning on the rfc have already been implemented in the past.
 
Event Description
It was reported that the error message "head" occured during service.No patient involvement.Internal reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTAFLOW CENTRIFUGAL PUMP SYSTEM
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key7789465
MDR Text Key117569039
Report Number8010762-2018-00246
Device Sequence Number1
Product Code DTQ
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROTAFLOW CONSOLE
Device Catalogue Number701043292
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/25/2018
Initial Date FDA Received08/16/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-