• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NEUROPACE, INC. NEUROPACE RNS SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEUROPACE, INC. NEUROPACE RNS SYSTEM Back to Search Results
Model Number DL-344-10-K
Device Problems Break (1069); High impedance (1291)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/12/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).The ecog and impedance data are suggestive of a potential lead break.Neuropace was unable to determine a root cause.The explanted product was not returned to neuropace, inc.
 
Event Description
On (b)(6) 2017 a signal artifact was observed on the left depth lead indicating a potential lead break.On (b)(6) 2017 a lead revision was performed and the depth lead with the signal artifact was explanted and a new lead was implanted.The explanted lead was not returned to neuropace for investigation.The patient does occasionally fall during seizures but the treating center was unable to determine if this was the cause of the possible break.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Type of Device
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer (Section G)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
Manufacturer Contact
ramona gonis
455 n. bernardo ave.
mountain view, CA 94043
6502382788
MDR Report Key7791583
MDR Text Key117419596
Report Number3004426659-2018-00037
Device Sequence Number1
Product Code PFN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/15/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model NumberDL-344-10-K
Device Catalogue Number1007440
Device Lot Number18860-1-1-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received12/12/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age36 YR
-
-