MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
|
Back to Search Results |
|
Model Number BI-700-00027-120 |
Device Problems
Mechanical Problem (1384); Failure to Fire (2610); Noise, Audible (3273)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/24/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Patient information not provided due to japanese patient privacy regulations.A medtronic representative went to the site to test the equipment.Testing revealed that after the standalone board and x relay were replaced, the system functioned as intended.The imaging system then passed the system checkout and was found to be fully functional.Missing device manufacture date.
|
|
Event Description
|
Medtronic received information regarding an imaging system being used during a sacroiliac and thoracolumbar procedure.It was reported that during the operation, the imaging system and navigation were used.Three (3)d image was taken and transferred normally twice.In the end, in order to confirm the setting condition of the screws, when an attempt was made to take 3d image, stand alone mode was indicated on the panel of image acquisition system (ias).The central x-ray part was showed as indicator.When the image acquisition system (ias) and mobile viewing station (mvs) were rebooted, an abnormal noise sounded like ¿gee¿ was heard and the x-ray indicator was not in ¿on¿ status.The use of the imaging system was aborted and the procedure was completed with no issue.This occurred intra/peri-operatively with no delay to surgical time.There was no reported impact to patient outcome.
|
|
Manufacturer Narrative
|
The standalone pcb board was returned to the manufacturer for analysis.Analysis found that the issue could be recreated.The standalone pcb board passed visual inspection, but failed bench testing.Standalone pcb powers up, all ac / dc voltages are present, both the fans are on.The 380vdc, 15vdc, and all leds are on.However, the vac is measured at 68vac while it suppose to measure at a round 110vac.Relay passed performance test.Varistor measured open, and both returned fuses are good.Standalone board failed due to low vac reading.Analysis found that the reported event was related to an electrical issue.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|