Model Number EK016R
Device Problems Structural Problem (2506); Sharp Edges (4013)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation results: date of event is unknown.The reported complaint referenced multiple aesculap trocars; however, not all referenced products were returned for investigation.Two trocars of the same lot (ek017r / lot 52415436) and one karl storz manufactured shaft were returned for investigation.The returned trocars were in good condition with no deviations noted by visual inspection.There were several scratches noted on the insulation of the karl storz shaft.In addition to the returned instruments, pictures were provided to assist in the investigation.The provided insulation pipe does not correspond to the provided pictures.The damage seen in the photo (a chip in the insulation) was not located on the provided insulation pipe; however, scratches were recognized.Indications of wear and tear were found on the inside and the outside of the distal ends of both returned trocars.The device quality and manufacturing history records have been reviewed for all available lot numbers.The device history records were found to be according to specification valid at the time of production.One similar incident from the same end user has been filed with products from the reported batch.Based on the investigation of the returned devices as well as review of all available information indicates, the reported failure is related to insufficient use of the device.The distal ends of the provided trocars were worn, but still according to specification.The damage on the shaft of the returned device, as well as the damage noted in the photos provided, can only occur when high lateral forces are applied on the shaft during movement of the trocar.A worn or damaged distal end of the trocar can enhance this effect.The reported failure mode is being reviewed and revised, as appropriate, by r&d.At the time of this report, no corrective/preventive action has been identified.
 
Event Description
Country of complaint: (b)(6).It was reported that a 5 mm instrument (karl storz manufactured shaft) sheared during use with an aesculap manufactured trocar.According to information available at the time of this report, there was negative consequence to the patient.Seven different item numbers of trocars were reported.It is not clear which of these items was in use when the insulation sheared; therefore, one medwatch report is being submitted for each reported item number.All products and medwatch reports related to this report are: ek011r / 9610612-2018-00354; ek016r / 9610612-2018-00355; ek017r / 9610612-2018-00356; ek018r / 9610612-2018-00357; ek024r / 9610612-2018-00358; ek028r / 9610612-2018-00359; ek036r / 9610612-2018-00360.Karl storz shaft (item number unknown) / included as concomitant product in each of the above medwatch reports.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TROCAR SLEEVE 5/110MM THREADED WITH TAP
Type of Device
TROCAR
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
nicole broyles
615 lambert pointe drive
hazelwood, MO 63042
MDR Report Key7795324
MDR Text Key117580057
Report Number9610612-2018-00355
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K101937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial
Report Date 07/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/17/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberEK016R
Device Catalogue NumberEK016R
Was Device Available for Evaluation? No
Distributor Facility Aware Date07/25/2018
Date Manufacturer Received07/19/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-