• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO PATIENT ROOM WALL SAVER RECL; CHAIR, EXAMINATION AND TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO PATIENT ROOM WALL SAVER RECL; CHAIR, EXAMINATION AND TREATMENT Back to Search Results
Catalog Number 3500000730
Device Problems Material Twisted/Bent (2981); Appropriate Term/Code Not Available (3191)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/19/2018
Event Type  malfunction  
Event Description
It was reported that the wall saver mechanisms have been bending, so that the latch will not hold and the recliner "pops open." no patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Manufacturer Narrative
Upon evaluation of the unit, the stryker service technician found that the footrest of the unit was stuck extended, and could not be lowered.Therefore, the alleged issue of the footrest not latching and popping open was not confirmed.The issue of a stuck footrest, unable to be lowered, is not likely to cause or contribute to serious injury or death if it was to recur and is, therefore, not reportable.
 
Event Description
It was reported that the wall saver mechanisms have been bending, so that the latch will not hold and the recliner "pops open." no patient was affected and no adverse consequence or clinically relevant delay in treatment was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT ROOM WALL SAVER RECL
Type of Device
CHAIR, EXAMINATION AND TREATMENT
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
mary klaver
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key7798203
MDR Text Key117848308
Report Number0001831750-2018-00980
Device Sequence Number1
Product Code FRK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/20/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Catalogue Number3500000730
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received07/25/2018
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/11/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-