Model Number 1000096 |
Device Problems
Disconnection (1171); Use of Device Problem (1670)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/29/2018 |
Event Type
malfunction
|
Manufacturer Narrative
|
Root cause: use error.Per tandem¿s t:slim x2g5 user guide: ¿check that your connection between the cartridge tubing and the infusion set tubing is tight and secure.¿ no product was returned for evaluation.Should new relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that the luer lock became disconnected.The customer's blood glucose level was 130 mg/dl.Reportedly, the customer replaced the infusion set.
|
|
Event Description
|
It was reported that the customer was new to the pump an did not fully tighten the tubing to the cartridge tubing resulting in the luer lock disconnecting.
|
|