Model Number 1458Q/86
Device Problems Incorrect Measurement (1383); Misconnection (1399); Separation Failure (2547)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/06/2018
Event Type  malfunction  
Event Description
During implant, the guidewire could not be inserted all the way through the lead.In addition, the pacing system analyzer (psa) was unable to correctly measure the electrical values due to a suspected issue with the connector at the top of the lead.A new lead was used to finish the procedure with no adverse consequences to the patient.
 
Manufacturer Narrative
As received, a complete lead without a guidewire was returned in one piece for analysis.A guidewire could not be inserted in lead due to blood clogged found inside the inner coil.Visual and x-ray inspections of the lead did not find any anomalies.The reported event of guidewire insertion failure was confirmed; however, the connector issue was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUARTET
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.
15900 valley view court
sylmar CA 91342
Manufacturer Contact
robert greenleaf
15900 valley view court
sylmar, CA 91342
8184932577
MDR Report Key7800848
MDR Text Key117725598
Report Number2017865-2018-12261
Device Sequence Number1
Product Code OJX
UDI-Device Identifier05414734503198
UDI-Public05414734503198
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/18/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/21/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/28/2021
Device Model Number1458Q/86
Device Lot NumberS000057976
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2018
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received09/17/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/19/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-