On (b)(6) 2018, the treating center contacted neuropace field support personnel after noting artifact on the ecog signal from the patient's right depth lead.After review by neuropace, inc., it was determined the signal artifact noted was consistent with those seen with a potential lead break.The lead break was attributed by the treating center , as a lead routing issue, as the lead had ended up folding over on itself.The lead was explanted on (b)(6) 2018 and a new lead was placed on (b)(6) 2018.The lead was discarded at the treating center per the neurosurgeon's instructions; and therefore was not returned to neuropace for analysis.
|