DEPUY IRELAND 9616671 ATTUNE IMPACTION HANDLE; ATTUNE INSTRUMENTS : HANDLES Back to Search Results
Catalog Number 254401017
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/26/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
During a total knee replacement the surgeon was impacting the femoral implant.On the third strike the impactor handle broke and the latch cracked off of the handle.All pieces were accounted for and the surgeon used the alternate impactor in the room.Since there was a second handle available there was no surgery delay or patient consequences.In the evening of the same day a sterile processing tech noticed cracks in a t handle, the cracks make compromise the sterility of the product and need to be replaced.(b)(4).Patient consequence? : no.Is the information being submitted for this complaint all the details that are known/available regarding this event? : yes.
 
Manufacturer Narrative
Investigation summary: the reported event has been evaluated and will be monitored.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE IMPACTION HANDLE
Type of Device
ATTUNE INSTRUMENTS : HANDLES
Manufacturer (Section D)
DEPUY IRELAND 9616671
loughbeg, ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910 
700 orthopaedic drive
warsaw IN 46582 0988
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46582-0988
5743725905
MDR Report Key7808346
MDR Text Key117960582
Report Number1818910-2018-67451
Device Sequence Number1
Product Code NJL
UDI-Device Identifier10603295423324
UDI-Public10603295423324
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 07/27/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254401017
Device Lot NumberNW156419
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received10/18/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/22/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-