• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND RESOLUTE ONYX RX; STENT, CORONARY, DRUG-ELUTING Back to Search Results
Catalog Number RONYX30038UX
Device Problems Migration or Expulsion of Device (1395); Difficult to Remove (1528); Activation, Positioning or Separation Problem (2906); Detachment of Device or Device Component (2907); Device Dislodged or Dislocated (2923)
Patient Problem Patient Problem/Medical Problem (2688)
Event Date 08/14/2018
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
During the procedure, an attempt was made to use one resolute onyx drug eluting stent to treat a mildly tortuous and mildly calcified lesion located in the mid right coronary artery, exhibiting 80% stenosis.There were no abnormalities reported in relation to anatomy.There was no damage noted to packaging.There were no issues noted when removing the device from the hoop.The device was inspected with no issues identified.Negative prep was performed with no issues noted.The lesion was pre-dilated.The device did not pass through a previously-deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.The resolute onyx balloon was inflated to 8atm and an attempt was then made to pull out the non-medtronic buddy wire.There were 2 wires placed down the rca.Physician was unable to remove the non-medtronic wire at 8 atm and inflated the stent balloon up to 16 atm for the full deployment of the stent.The non-medtronic wires were unable to be removed after the second inflation.Physician then tried to remove the entire system and noted that the balloon shaft had been snapped in half distal to the balloon.The device was not kinked and re-straightened during use.The portion of the device does not remain in patient.The remaining part to the stent catheter was removed using a snare.Patient status post-procedure is alive with no injury.
 
Manufacturer Narrative
Additional information: the stent was not completed deployed, and the stent came off the catheter, and ended up migrating down the rca.The stent was retrieved from the rca by a snare technique and was completely removed out of the rca.It is unknown what happened first, the shaft breaking or the stent coming off.Evaluation summary: the stent was not present on the balloon and did not return for analysis.The balloon folds were expanded.Crimp impressions were visible on the exposed balloon surface.The inner lumen could not be verified with a 0.015 inch mandrel most likely due to hardened blood in the guidewire lumen.The delivery system returned with a detachment on the transitional tubing 35.3cm proximal to the distal tip.The transitional tubing material was oval and jagged on both sides of the detachment site.Kinks were not evident on the transitional tubing proximal or distal to the detachment site.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Type of Device
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
Manufacturer Contact
toni o'doherty
parkmore business park west
galway 
091708734
MDR Report Key7812213
MDR Text Key118086610
Report Number9612164-2018-02158
Device Sequence Number1
Product Code NIQ
UDI-Device Identifier00643169557048
UDI-Public00643169557048
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P160043
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 12/03/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/24/2018
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/02/2020
Device Catalogue NumberRONYX30038UX
Device Lot Number0009016962
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2018
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/06/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/03/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age72 YR
-
-