Model Number 595000-001
Device Problem Audible Prompt/Feedback Problem (4020)
Patient Problems High Blood Pressure/ Hypertension (1908); No Consequences Or Impact To Patient (2199)
Event Date 08/21/2018
Event Type  malfunction  
Manufacturer Narrative
This alleged failure mode poses a low risk to the patient because although the freedom driver exhibited a fault alarm, it continued to perform its life-sustaining functions.The freedom driver will be returned to syncardia for evaluation.The results of the investigation will be provided in a follow-up mdr.
 
Event Description
The customer, a syncardia authorized distributor, reported that the freedom driver exhibited a fault alarm.The customer also reported that the patient was hypertensive.The patient was subsequently switched to a backup freedom driver without any adverse patient impact.
 
Manufacturer Narrative
Visual inspection identified a broken external connector on the connection 1 receptacle cable.The root cause of the damaged external connector cannot be conclusively determined, but is likely the result of excessive extraction force being applied with disconnecting the power adaptor from the driver, or as a result of rough handling/impact shock.The driver's alarm history was reviewed and revealed a '34' fault code alarm which could have been the customer-reported alarm.This type of alarm is usually produced because of intermittent power communication errors and it correlates with the broken external connector of connection 1 receptacle cable.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation and is closing this file.(b)(4) follow-up report 1.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNCARDIA FREEDOM DRIVER
Type of Device
EXTERNAL PNEUMATIC DRIVER
Manufacturer (Section D)
SYNCARDIA SYSTEMS, LLC
1992 e. silverlake road
tucson AZ 85713
MDR Report Key7822473
MDR Text Key118550571
Report Number3003761017-2018-00450
Device Sequence Number1
Product Code LOZ
UDI-Device Identifier00858000003121
UDI-Public(01)00858000003121
Combination Product (y/n)N
PMA/PMN Number
P030011
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign
Type of Report Initial,Followup
Report Date 12/06/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number595000-001
Device Catalogue Number595000-001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/29/2018
Date Manufacturer Received08/21/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age42 YR
-
-