Model Number BI-700-00027-120
Device Problem Failure to Fire (2610)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/03/2018
Event Type  malfunction  
Manufacturer Narrative
Patient information was unavailable from the site.Missing udi.A medtronic representative went to the site to test the equipment.Testing revealed that the system was functioning as intended.The system passed the system checkout and was found to be fully functional.
 
Event Description
Medtronic received information regarding an imaging system.It was reported that the gantry was not "spinning".There was no patient impact reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
maxwell roth
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
MDR Report Key7823915
MDR Text Key118533531
Report Number1723170-2018-04275
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/28/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/28/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/03/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/15/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-