It was reported that there was a hole in the catheter.An angiojet solent omni catheter was selected for use during a thrombectomy procedure.The device primed successfully, but when attempting to use the device in the patient, a hole was noticed on the catheter and the device was not functional.It is not known what caused the hole.The catheter was removed replaced with another of the same device.No complications were reported.The patient's condition after the procedure was fine.The returned device consisted of a solent omni thrombectomy system.The pump assembly, effluent/supply line, shaft and tip were microscopically and visually inspected.Blood was present outside and inside the device and there was 200ml of blood in the attached waste bag, when received.Visual inspection revealed that there was damage to the shaft in the proximal end of the catheter.Microscopic examination revealed that there was a hole in the pebax shaft 53.1cm to 53.5cm distal of the strain relief.Functional testing was performed by placing the device in the angiojet ultra console.Functional testing consisted of running the complaint device through the full priming cycle in thrombectomy mode; however, the device leaked from the hole in the shaft.Inspection of the remainder of the device presented no other damage or irregularities.
|