• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM O2 IMAGING SYSTEM; INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM Back to Search Results
Catalog Number BI70002000-G27
Device Problem Difficult to Open or Close (2921)
Patient Problem No Patient Involvement (2645)
Event Date 08/07/2018
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Udi and manufacturing date not available for this system at time of filing.Onsite functional and visual examination was performed by a manufacturer representative.The systems movement mechanisms were adjusted and the issue was resolved.Onsite functional and visual examination was performed by a manufacturer representative.
 
Event Description
Medtronic received information regarding an imaging device being used outside of a procedure.It was reported that while setting up for a lab today the system looks to be shut, but the system states it's open.He also stated the rotor will not move.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM O2 IMAGING SYSTEM
Type of Device
INTERVENTIONAL FLUOROSCOPIC X-RAY SYSTEM
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
parker desautel
navigation customer quality
826 coal creek circle
louisville, CO 80027
MDR Report Key7840110
MDR Text Key119032858
Report Number1723170-2018-04355
Device Sequence Number1
Product Code OWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 08/31/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/31/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBI70002000-G27
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/07/2018
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-