It was reported that the stent was shorter than the labeled size.Two 7x80x130 innova self expanding stents were selected for use for a stenting procedure in the right iliac artery.The first stent was deployed successfully and had a length of 80mm.The second stent was also deployed successfully.Balloon angioplasty was done after both stents were deployed.Upon radiographic imaging, it appeared that the second stent was significantly shorter, around 60mm fully deployed.The stent still covered the entire lesion.The case was completed with the same devices.There were no patient complications reported.Device evaluated by mfr: the outer shaft, middle shaft, inner sheath and the remainder of the device were checked for damage.The outer sheath was kinked at the nosecone.The inner liner showed a kink at 9.5cm from the tip.The stent was not returned; therefore, the length of the stent could not be measured.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.
|