Catalog Number 35302
Device Problems Positioning Failure (1158); Nonstandard Device (1420)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The device was not returned.
 
Event Description
It was reported that the sheaths were not staying on and allegedly caused the patient distress due to wetness.The patient also had problems with the sheaths not unrolling properly.
 
Event Description
It was reported that the sheaths were not staying on and allegedly caused the patient distress due to wetness.The patient also had problems with the sheaths not unrolling properly.
 
Manufacturer Narrative
The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use was found adequate and states the following: "fitting the sheath: the sheath is a strapless penile sheath: it has an adhesive coating inside to ensure a safe and simple procedure for fitting.If necessary trim pubic hair clean and dry pubic area remove sheath from its wrapping place rolled end over the end of the penis, leaving a small space between the end of the penis and the cup of the sheath if uncircumcised, do not retract the foreskin but allow it to remain over the glans.Avoid contact between the adhesive and the glans.Slowly unroll the sheath along the shaft of the penis, then gently squeeze the shad around the penis to ensure even adhesion.Try to avoid leaving a rolled 'collar' of sheath around the base of the penis.Finally, connect the urine collection bag.Always check that the connections are secure before use.Wear time will vary from user to user.It is recommended that a sheath b changed every 24 hours for reasons of hygiene.To remove : the sheath may be removed by gently tolling it off the penis.Avoid simply lulling the sheath off.Removing the sheath whilst having a bath or shower may increase comfort.Do not use on irritated or compromised skin.Discontinue use if irritation occurs." section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC SPIRIT SHEATH STYLE 1 STANDARD 29MM 30BX
Type of Device
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7864487
MDR Text Key120105338
Report Number1018233-2018-04108
Device Sequence Number1
Product Code NNX
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 09/25/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/11/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/28/2023
Device Catalogue Number35302
Device Lot NumberJUCR0309
Was Device Available for Evaluation? No
Date Manufacturer Received09/24/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-