• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Model Number 0165SI14
Device Problems Inaccurate Flow Rate (1249); No Flow (2991)
Patient Problems No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Event Date 08/15/2018
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event Description
It was reported that there was no urine flow after placement.The catheter was replaced and urine flowed out.
 
Event Description
It was reported that there was no urine flow after placement.The catheter was replaced and urine flowed out.
 
Manufacturer Narrative
The reported issue was unconfirmed.The device was returned and visually inspected.No visual defects were noted on returned catheter.Evaluation found no blockage in the drainage lumen.Water was introduced through the drainage eye and came out from drainage funnel without difficulty.Sample was sent for flow rate test and passed.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿5) do not aspirate urine through drainage funnel wall.6) since movement of the body, etc.May twist or bend catheter to cause occlusion, care should be taken to fix the catheter securely.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key7868532
MDR Text Key120127091
Report Number1018233-2018-04138
Device Sequence Number1
Product Code MJC
UDI-Device Identifier00801741016745
UDI-Public(01)00801741016745
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 11/08/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/12/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/01/2021
Device Model Number0165SI14
Device Catalogue Number0165SI14
Device Lot Number8BM8146
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2018
Date Manufacturer Received10/17/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-