• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NATUS MEDICAL INCORPORATED OLYMPIC COOL-CAP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NATUS MEDICAL INCORPORATED OLYMPIC COOL-CAP Back to Search Results
Model Number 005094
Device Problem Insufficient Cooling (1130)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Natus technical service has requested return of the system.Natus technical service has sent questions to the complainant regarding the details of the event and the patient involved in the event.A response to these requests has not been received.
 
Event Description
It was reported that a cool-cap system's cap temperature was high as compared to the set temperature.A patient was being treated with the system when the event occurred, and the patient was moved to another cool-cap system.There was no death, serious injury, delay in treatment, or environmental/safety concerns.A service technician who reported the event performed a heat load test as described in the cool-cap service manual and functional test procedure.The technician claimed that the test demonstrated that the internal thermistors were faulty.However, based on photos provided by the technician, natus factory service noted that the heat load test was performed incorrectly.Natus factory service stated that testing the functionality of the cooling modules would also require performing a different test.
 
Manufacturer Narrative
Natus technical service has requested return of the system, the unit was received by natus for repair on 10/23/2018.Natus factory service confirmed customer complaint and found a faulty cooling module on b side of cooling block.Thus, only one side of the cooling block was functioning, this was the root cause of the issue.Both cooling modules inside cooling block were replaced and the problem was fixed.Cd drive was also found to have tray issue and was replaced.Unit was fixed and passed cool cap ftp doc-001692 rev.O.The unit was sent to customer on 11/08/2018 and natus technical service made follow up with customer to check the status of the device.Customer confirmed that the equipment has been successfully installed with no issues.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OLYMPIC COOL-CAP
Type of Device
COOL-CAP
Manufacturer (Section D)
NATUS MEDICAL INCORPORATED
5900 first avenue south
seattle WA 98108
MDR Report Key7873604
MDR Text Key120256986
Report Number3018859-2018-00150
Device Sequence Number1
Product Code MXM
Combination Product (y/n)N
PMA/PMN Number
P040025
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Remedial Action Repair
Type of Report Initial,Followup
Report Date 08/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/13/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model Number005094
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2018
Date Manufacturer Received10/23/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-