• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CV BOM 7MM EXTENDED LENGTH ENDOSCOPE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CV BOM 7MM EXTENDED LENGTH ENDOSCOPE; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Catalog Number C-VH-1111
Device Problems Structural Problem (2506); Material Deformation (2976)
Patient Problem No Patient Involvement (2645)
Event Date 08/23/2018
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Since the device is not available to be returned to us, a technical evaluation cannot be performed.Per our standard sop's, all events are tracked and trended to determine whether or not any trends develop.
 
Event Description
The hospital reported that during sterilization/sanitization for an endoscopic vein harvesting procedure using bom 7mm extended length endoscope.It was stated that they used an nx sterrad model to sterilize their vh-1111 endoscope.It was also stated "it is causing scopes to fail and becoming physically compromised".The proximal end of the scope looks as if it is "melting".It was made aware that the nx sterrad model is not listed on getinge instructions for sterilization for vh-1111.No patient involvement.
 
Manufacturer Narrative
Corrected section: patient code changed to "no patient involvement" internal complaint number: tw# (b)(4).Autonumber: (b)(4).This is a reusable oem device; therefore, a lot history review was not applicable.A lot/serial number was not provided and the specific product lot/serial number cannot be identified from a ship history, therefore it is impossible to obtain a certificate of conformance.
 
Event Description
The hospital reported that during sterilization/sanitization for an endoscopic vein harvesting procedure using bom 7mm extended length endoscope.It was stated that they used an nx sterrad model to sterilize their vh-1111 endoscope.It was also stated "it is causing scopes to fail and becoming physically compromised".The proximal end of the scope looks as if it is "melting".It was made aware that the nx sterrad model is not listed on getinge instructions for sterilization for vh-1111.No patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BOM 7MM EXTENDED LENGTH ENDOSCOPE
Type of Device
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
MAQUET CV
45 barbour pond drive
wayne NJ 07470
MDR Report Key7878702
MDR Text Key120529161
Report Number2242352-2018-00906
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
PMA/PMN Number
K014250
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/14/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/14/2018
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberC-VH-1111
Was Device Available for Evaluation? No
Date Manufacturer Received10/15/2018
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-